The compositive tester uses 12digit high-speed A/D converter. The speed reaches 10MHz when sampling. It can test the wave shape of a single high frequency wave. Also, it provides reliable methods for the researchers to develop the ballast working principle and analyze reliably. It meets the ballast and fluorescent lamp as well as the electrical capability testing requirements of IEC60929,IEC60969,IEC61000-3-2.
Measure parameters
Input: Vrms,Irms, W, PF, Hz, phase angle, total harmonic and 0~ 39th component of harmonic
Output:Lamp voltage, lamp current, filament current, guided cathodic current, lamp power, crest factor and oscillatory frequency, single high frequence pulse analysis
Startup:Startup lamp voltage, lamp current, filament current, guided cathodic current curve and parameters, preheating time
> Input characteristic test
a) Test voltage, current, power, power factor, power line frequency, total harmonic as well as 0~50 time harmonic separate amount and phrase angle
b) Range of voltage and current basic wave frequency:45Hz-65Hz
Range of narrow frequency: 45Hz-5kHz
Range of wide frequency: 45Hz-1MHz
c) Range of voltage: 10.0-300.0V (virtual value)
d) Range of current: NR: 0.010-1.500A (virtual value)
WR: 0.010-4.500A (virtual value)
e) Range of power: NR: 0-450.0W
WR: 0-999.9W
Range of power factor: 0.000-1.000
> Output stable test
a) Test lamp voltage, current, power, filament current, guide cathode current, wave crest coefficient and surge frequency
b) Range of lamp voltage: 10.0-300.0V
c) Range of lamp current, filament current and guide cathode current
NR: 0.010-0.750A
WR: 0.010-1.500A
d) Range of lamp power: NR: 0.5-200.0W
WR: 0.5-400.0W
> Output startup test
a) Test preheat time and lamp voltage, current, filament current, changing curve and data of guide cathode current within 0 to 5s
b) Range of lamp voltage: 10.0-800.0V
c) Range of lamp current, filament current and guide cathode current
NR: 0.010-0.750A
WR: 0.010-1.500A